Our growth depends to the superior products ,great talents and repeatedly strengthened technology forces for Animal Liver Test Kit , Unisensor Test Kit , Wine Test kit , Welcome your visiting and any your inquires,sincerely hope we can have chance to cooperate with you and we can build up long well business relationship with you.
2021 Good Quality SulfaSensor Test Kit - HoneyGuard Tetracyclines Test Kit – kwinbon Detail:

Cat. KB01009K-50T

About
This kit is used for rapid qualitative analysis of tetracyclines in honey sample.

Sample prepare method
(1) If the honey sample crystallized ,warm it in water bath not high than 60℃, till the honey sample thaw ,mix completely ,cooling as room temperature , then weight for assay.
(2) Weigh 1.0±0.05g homogenate into a 10ml polystyrene centrifuge tube, add 3ml sample extraction solution, vortex for 2min or shake it by hand until the sample is mixed completely.

Assay operations.
(1.) Take bottles needed from the kit package, take out required cards, and make proper marks. Please use these test cards within 1h after open package.
(2.) Take 100l prepared sample into the sample hole by pipette, then start the timer after liquid flow.
(3.) Incubate for 10min at room temperature.
7.LOD

Tetracyclines

LOD(μg/L)

Tetracyclines

LOD(μg/L)

tetracycline

10

doxycycline 15
aureomycin

20

oxytetracycline

10

Results
There are 2 lines in the card result area, Control line and Tetracylcines Line, which are briefly  sed as “B” and “T”. The test results will depend on the color of these lines. The following diagram describes the result identification.
Negative: Control line and Test line both are red and T Line is darker than the control line;
Tetracyclines Positive: Control Line is red, T Line has no color or T Line is lighter color than C line, or the T Line is the same as C Line.

Results

Storage
2-30°C in dark dry place, do not freeze. The kit will be valid in 12 months. The lot number and expired date are printed on the package.


Product detail pictures:

2021 Good Quality SulfaSensor Test Kit - HoneyGuard Tetracyclines Test Kit – kwinbon detail pictures

2021 Good Quality SulfaSensor Test Kit - HoneyGuard Tetracyclines Test Kit – kwinbon detail pictures

2021 Good Quality SulfaSensor Test Kit - HoneyGuard Tetracyclines Test Kit – kwinbon detail pictures


Related Product Guide:

Our company has been focusing on brand strategy. Customers' satisfaction is our best advertising. We also supply OEM service for 2021 Good Quality SulfaSensor Test Kit - HoneyGuard Tetracyclines Test Kit – kwinbon , The product will supply to all over the world, such as: Croatia, Victoria, Greece, We are committed to meet all your needs and solve any technical problems you may encounter with your industrial components. Our exceptional products and vast knowledge of technology makes us the preferred choice for our customers.
  • Hope that the company could stick to the enterprise spirit of "Quality, Efficiency, Innovation and Integrity", it will be better and better in the future.
    5 Stars By Darlene from Gambia - 2018.06.12 16:22
    We have been engaged in this industry for many years, we appreciate the work attitude and production capacity of the company, this is a reputable and professional manufacturer.
    5 Stars By Joseph from Oman - 2018.06.18 19:26
    Write your message here and send it to us